In the base model, total carbon (TC) along with total organic carbon (TOC) and inorganic carbon (IC) in water/liquid samples are analyzed by utilizing high temperature catalytic combustion technique. Total nitrogen parameter is an optional addition. Proven chemiluminescence technique allows low detection limits.
TRL-TOC-L-C Series is a NEW MODEL with built in liquid handling module automating sparging for direct TOC and IC analysis. Sample loading automation is possible by choosing one of the optional autosamplers.
TRL-TOC-L-C Series comply with most international standards including
ISO8245, ASTM D7573-18, Method 9060A, EPA415.1, Standard Method 5310B, EN13137
Method of analysis | High Temperature Catalytic Oxidation |
Carbon Detection | NDIR |
Nitrogen Analysis | with optional TRL-TN |
Nitrogen Detection | Chemiluminescence |
LOD for Carbon in gr | 0.25ug |
LOD for Nitrogen gr | 0.1ug |
LOD for Carbon in concentration | 4-50ppb depending on carrier gas quality |
LOD for Nitrogen concentration | 20ppb |
Analysis range for Carbon | 60000ppm |
Analysis range for Nitrogen | 10000ppm |
Type of Sample | Liquid |
Analyzed Parameters | TC, IC, TOC(NPOC), Optional TN |
Caculated Parameters | None |
Furnace Temperature | Adjustable dual furnace, operating between 450-1100°C |
Sample Loading for TC/TN/TOC Analysis | In 40mL vials |
Sample Loading for IC Analysis | the same vial is for all parameters |
Sample Loading Automation | With optional TRL-LSA unit |
Manual Sample Loading | Sample suction from sample cup |
IC Analysis | By selecting through s/w |
Sample Dilution | By selecting through s/w |
Flow Modes | Single |
Carrier Gas Flow Control | By Mass Flow Control Valve driven via s/w |
Carrier Gas Conditions | 100-300mL/min, oeprator adjustable, MFC controlled |
Operating Conditions | 5 to 45°C |
Power | 1100VA, 230VAC |
Dimensions, Analyzer HxWxD | 65x59x63 cm |