The Testomat® 808 SiO2 has been designed for use in the sterilisation of equipment in hospitals. It complies with the EN 285:2006 standard. The device is a limit gauge that automatically monitors the level of SiO2 in the water. It gives you the choice between 10 limits in the range of 0.3 to 1.2 ppm.
Silica SiO2: 0.3 – 1.2 ppm
• Measuring chamber flushed with pressure FOR SILICA UP TO 1.2 PPM
• Low water consumption
• Cutting-edge electronics
• State-of-the-art indicator pump system
• Direct error and indicator quantity display
• Quality management with 2 relay outputs
• Limit value evaluation / External control
• Alarm processing
• Internal flushing via manual control
• 72 hours of unsupervised operation possible
• Selection between two indicator bottle sizes
• Two selector switches for interval measurement
and limit value evaluation
Reagents | Art. no. 100 ml bottle | Art. no. 500 ml bottle |
Testomat® 808 SiO2 reagent set, reagents A + B | 140808 | – |
Testomat® 808 SiO2 reagent A | – | 141808 |
Testomat® 808 SiO2 reagent B | – | 141808 |
The Testomat® 808 SiO2 equipment has been
designed for use in the sterilisation of hospitals. The device is a limit gauge that automatically monitors the level of SiO2
in the water. It
complies with the EN 285:2006 standard for
steam sterilizers.
The device is also suited in other industries for
product monitoring due to its adjustable measuring range.
• Selection of 10 limit values from 0.3 to
1.2 ppm
• Automatic interval mode
interval pause can be set from
0 – 480 minutes
• External control (quit alarms,
stop analysis)
• Manual start
• Extended operating periods due to 500 ml
indicator storage bottle
• RS232 interface for optional firmware update
and data output to the computer
• Weekend operation monitoring through 72-
hour operation without supervision (BOB)
• Status and error messages output via a
current loop